1529 Chub-E4 Standards Thermometer

Feature Summary
- Four channels for PRTs, thermistors, and thermocouples
- Displays eight user-selected data fields from any channel
- Logs up to 8,000 readings with date and time stamps
- Battery provides eight hours of continuous operation
Specifications
|
PRT / RTD |
Thermistor |
Thermocouple |
Inputs |
2
channels PRT/thermistor and 2 channels TC, or 4 channels PRT/thermistor, or 4 channels TC,
specify when ordering; PRT/thermistor channels accept 2, 3, or 4 wires; TC inputs accept
B, E, J, K, N, R, S, T, and Au-Pt TC types |
Temperature Range |
189 to 960°C |
50 to 150°C |
270 to 1800°C
|
Measurement Range |
0 to 400 W |
0 to 500 KW |
10 to 100 mV |
Characterizations |
ITS-90, IEC-751 (DIN
385), Callendar-Van Dusen |
Steinhart-Hart,
YSI-400 |
NIST Monograph 175,
3-point deviation function applied to NIST 175, 6th-order polynomial |
Temperature Accuracy
(meter only) |
±0.004°C at
100°C
±0.006°C at 0°C
±0.009°C at 100°C
±0.012°C at 200°C
±0.018°C at 400°C
±0.024°C at 600°C |
±0.0025°C at 0°C
±0.0025°C at 25°C
±0.004°C at 50°C
±0.010°C at 75°C
±0.025°C at 100°C |
Ext.
RJC Int. RJC
B at
1000°C ±0.6°C ±0.6°C
E at
600°C ±0.07°C ±0.25°C
J at
600°C ±0.1°C ±0.35°C
K at
600°C ±0.15°C ±0.4°C
N at
600°C ±0.15°C ±0.3°C
R at
1000°C ±0.4°C ±0.5°C
S at
1000°C ±0.5°C ±0.6°C
T at
200°C ±0.1°C ±0.3°C |
Temperature Resolution
|
0.001° |
0.0001° |
0.01 to 0.001° |
Resistance / Voltage
Accuracy |
0 to 20 W: ±0.0005 W
20 to 400 W: ±25 ppm of rdg. |
0 to 5 KW: ±0.5 W
5 to 200 KW: ±100 ppm of rdg.
200 to 500 KW ±300 ppm of rdg. |
10 to 50 mV:
±0.005 mV
50 to 100 mV: ±100 ppm of rdg.
(Internal RJC: ±0.25°C) |
Operating Range |
16 to
30°C |
Measurement Interval
|
0.1
second to 1 hour; inputs may be read sequentially or simultaneously at 1 second or greater
interval |
Excitation Current |
1mA |
2 and 10 mA, automatically
selected |
n/a |
Display |
1.3
x 5 backlit LCD graphical display |
Display Units |
°C,
°F, K, W, KW, mV |
Data Logging |
Up to
8,000 time- and date-stamped measurements can be logged |
Logging Intervals |
0.1,
0.2, 0.5, 1, 2, 5, 10, 30, or 60 seconds; 2, 5, 10, 30, or 60 minutes |
Averaging |
Moving
average of most recent 2 to 10 readings, user selectable |
Probe Connection |
Patented
DWF Connectors accept spade lug, bare-wire, or banana plug terminations |
Universal receptacle
accepts miniature and standard connectors |
Communications |
RS-232
and IR ports included, IEEE-488 (GPIB) optional |
AC Power |
100 to
240V AC, 50-60 Hz, 0.4A |
DC Power |
12 to
16V DC, 0.5A (battery charges during operation from 14.5 to 16V DC, 1.0A) |
Battery |
NiMH, 8
hours operation typical without backlight, 3 hours to charge, 500 cycles |
Size |
4.0
H x 7.5 W x 8.2 D (102 x 191 x 208 mm) |
Weight |
4.5
lbs. (2 kg) |
Probes from Hart |
Contact
Hart for a wide variety of precision PRTs, thermistors, and thermocouples |
Calibration |
NIST-traceable
resistance certificate and NIST-traceable voltage certificate provided |
So you need multiple channels, battery power, outstanding accuracy, and
the ability to read PRTs, thermistors, and thermocouples. Well, heres your answer.
Harts new Model 1529 Chub-E4 Thermometer gives you four channels,
three major sensor types, lab-quality accuracy, and a ton of great features, all at a
price youll love.
Inputs
The Chub-E4 has four inputs for reading four different sensors
simultaneously, and well configure those inputs in any of three different ways.
Choose four channels of thermocouple inputs, four channels of PRT/thermistor inputs, or
two channels of each. With this thermometer, reading thermocouples, PRTs, and thermistors
accurately from the same device is no problem.
100-ohm or 25-ohm PRTs (with two, three, or four wires) are read using
ITS-90, IEC-751 (DIN), or Callendar-Van Dusen conversion methods. Typical accuracies
include ±0.006°C at temperatures below 0°C and ±0.009°C at 100°C. Thermistor
readings are converted using the Steinhart-Hart polynomial or standard YSI-400 curve and
are as accurate as ±0.0025°C at 25°C with four-digit resolution.
Thermocouple inputs read all the common thermocouple types, including B,
E, J, K, N, R, S, T, and Au-Pt, and allow you to choose between internal and external
reference junction compensation. Typical accuracy for a type J thermocouple using internal
reference junction compensation at 600°C is ±0.35°C, not including the thermocouple.
PRTs and thermistors connect easily to the 1529 using Harts patented
mini-DWF connectors, which accept bare-wire, spade lug, or miniature banana plug
terminations. Thermocouples connect using standard or miniature terminations. Measurements
are made each second and can be taken simultaneously or sequentially. A special high-speed
mode allows measurements to be taken at the rate of 10 per second.
Display
If you think three sensor types and four inputs sounds versatile, wait
until you see the display panel on the Chub-E4. Displaying measurements in °C, °F, K,
ohms, or millivolts and choosing temperature resolution from 1 to 0.0001 are just the
beginning.
You can also select any eight data fields to view on-screen. Choose
statistical functions such as averages, standard deviations, and spreads; choose probe
information such as probe type and serial number; choose T1-T2 functions using inputs from
any two channels; or choose utility functions such as the date, time, and logging status.
You can even save a number of screen configurations for easy recall.
The push of a front-panel button also brings up a simple menu system to
easily guide you through the set-up and memory options of the 1529. Probe coefficients,
sample intervals, communication settings, password settings, and a host of other functions
are all easily accessible.
|
Choose from three combinations of inputs: 2 PRT/Thermistor and 2 TC or 4 PRT/Thermistor or 4 TC.
PRTs and thermistors connect easily with Hart’s patented mini-DWF connectors, which accept bare wire, spade lug, or banana plug terminations.
The Chub-E4 reads 2-, 3-, or 4-wire PRTs with either 25- or 100-ohm nominal resistance values. A grounding terminal is also included.
Thermocouple receptacles accept both standard and miniature connectors. The Chub-E4 reads thermocouple types B, E, J, K, N, R, S, T, and Au-PT.
|
Communications
The memory and communications capabilities of the Chub-E4 make it perfect
for benchtop thermometry, on-site measurements, lab calibration work, and remote data
logging. Optional software packages from Hart make this one of the most powerful
thermometers on the market.
With battery power and memory to store up to 8,000 measurements (including
date and time stamps), the 1529 has plenty of data logging capability. Store individual
measurements or any number of automatic log sessions (up to 8,000 readings), each tagged
with an identifying session label. Fourteen different logging intervals may be selected,
from 0.1 second to 60 minutes.
With Harts Model 9935 LogWare II, data may be quickly downloaded to
your PC for complete graphical and statistical analysis. Separate log sessions may be
automatically downloaded to separate files based on session labels. With this software,
the 1529 can even be used for real-time data logging. Log four channels at once directly
to your PC with virtually no limit to the number of data points you take. You can analyze
data, set alarm events, and even set delayed start and stop times.
With Calibrate-it software, the Chub-E4 may be integrated into a
completely automated calibration system. Use one input for your reference thermometer and
calibrate up to three other thermometers automatically. An RS-232 port is standard on
every unit and an IEEE-488 port is optional.
More Great Features
The 1529 runs on AC power from 100 to 240 volts, DC power from 12 to 16
volts, or off its internal nickel-metal-hydride battery for up to eight hours between
charging. The standard battery charges in less than three hours and lasts through 500
charge/recharge cycles.
If you want to rack mount your Chub-E4, weve even got a rack-mount
kit for you. This unit fits on your benchtop, in your instrument rack, or in your hand.
Of course, all the reference thermometers you might need for your 1529 are available from Hart, including secondary standard PRTs, standard thermistors, and noble-metal thermocouples. Carrying cases and even a serial printer for direct printer output are also available.
Weve said it before and well keep saying it: Hart Scientific
makes the best thermometer readouts in the world. No one else gives you a comparable
combination of accuracy, versatility, productivity enhancing features, and price.
Ordering information
| 1529 |
Chub-E4 Thermometer, 2 TC and 2 PRT/Thermistor inputs |
| 1529-R |
Chub-E4 Thermometer, 4 PRT/Thermistor inputs |
| 1529-T |
Chub-E4 Thermometer, 4 TC inputs |
| 2506-1529 |
IEEE Option |
| 9322 |
Rugged Carrying Case, holds 1529 and four probes up to 12” long |
| 2513-1529 |
Rack-Mount Kit
9935-S LogWare II, Multi Channel, Single User |
| 9935-M |
LogWare II, Multi Channel, Multi User |
| 2362 |
Spare AC Adapter, 15 V |
| 9320A |
External Battery Pack |
Downloads